Biological Safety Cabinet Class II B2 56-BSC203 Catalog

Download

Overview :

Biological Safety Cabinet Class II B2 56-BSC203 has safety level 2 enclosed cabinet unit where there is no recirculation inside the workspace; 100 % air is exhausted as they are hard connected to exhaust system. This is used to handle volatile toxic chemicals and radio nucleotides where chemical vapor re-circulation into the work zone is not permitted. It meets the requirements for the protection of product, personnel and the environment allowing for the safe handling of pathogens, contaminants or other potentially hazardous materials.

Features:

  • 100% HEPA-filtered exhaust airflow with zero internal recirculation
  • 0.3 µm HEPA filtration maintaining Level 100 clean environment
  • Ascending airflow ≥0.55 m/s with 0.35 m/s vertical laminar downflow
  • Vertical negative pressure system hard-ducted to facility exhaust

Applications:

FunctionImplementationHigh-volume toxic chemical handlingLarge fully exhausted workspace preventing vapor re-entryRadioactive tracer proceduresHard-connected exhaust system for radionuclide containmentPathogen manipulationCFU ≤10 personnel protection within validated airflow fieldSterile production preparationCFU ≤5 product protection for culture and media processingPharmaceutical and industrial labsExpanded work zone supporting larger apparatus and batch tasks

Specifications:
Air exhaust 100%
Glass door orientation 10° tilted
HEPA filter 0.3 micron
Air pressure Vertical negative
Cleaning level 100 level
Ascending wind speed ≥ 0.55 m/s
Descending wind speed ≥ 0.35 m/s
Noise ≤ 58 dB
Half ventilation peak value ≤ 3 µm
Personnel protection CFU ≤ 10
Product protection CFU ≤ 5
Cross contamination CFU ≤ 2
Fluorescent lamp 30 W x 2
UV lamp 30 W x 2
Power AC 220 V / 50 Hz
Air supply filter dimension 1590 x 554 x 50 mm
Working area 1600 x 650 x 620 mm
Overall dimension 1795 x 720 x 1950 mm
Weight 380 kg