Biological Safety Cabinet Class II A2 LE-BSC303

Biological Safety Cabinet Class II A2 LE-BSC303

Biological Safety Cabinet Class II A2 LE-BSC303 maintains 70% recirculated airflow and 30% exhaust, delivering a sterile and contamination-free workspace. It features a microprocessor system that stabilizes airflow without manual speed regulation. Built-in quiet motor with low energy consumption ensures minimal heat output and long-term reliability. Integrated lighting with concealed tubes reduces eye strain, while the interlocking UV safety system enhances operator protection. Our Biological Safety Cabinet Class II A2 is essential for contamination-sensitive procedures involving pathogenic or hazardous biological agents.

Specifications

Type of BSC Class Ⅱ Type A2
Airflow Mode 70% Cycle, 30% Exhaust
Exhaust Direction Top Exhaust
Average Inflow Velocity 0.50m/s
Average Downflow Velocity 0.33m/s
Inflow Airflow Volume 549m3/
Downflow Airflow Volume 1277m3/h
Exhaust Airflow Volume 549m3/h
Velocity Meter Accuracy 0.001m/s
Cleanliness ULPA: Class 3(ISO 14644.1); Class 100 (Fed 209E)
Filter Typical Efficiency Greater Than 99.999% at 0.1 to 0.3 Micron, ULPA as per IEST-RP-CC001.3 USA
Total Colony in Impaction Sampler (Operator Protection) ≤10CFU/Time
Total Colony in Slit Type Sampler (Operator Protection) ≤5CFU/Time
Total Colony in Culture Dish (Product Protection) ≤5CFU/Time
Total Colony in Culture Dish (Cross Contamination) ≤2CFU/Time
Personnel Protection Test KI-Discus Containment and Microbiological Testing are Performed
Blower AC
Downflow Filter ULPA
Exhaust Filter ULPA
Main Body 1.2mm (xxx'') 18 Gauge Electro-Galvanized Steel with White Oven-Baked Powder-Coated Finish
Work Zone 1.5mm (xxx'') 16 Gauge Stainless Steel, Type304, Single-Piece or 3-Piece
Side Walls 1.5mm (Xxx'') 16 Gauge Stainless Steel, Type304
Sash Glass Thickness 5mm
Sash Glass Type Tempered Glass, UV-Proof
Sash Window Working Opening 180mm
Sash Maximum Opening 530mm
Fluorescent Light Intensity ≥700lux
RMS ≤3um
Sound Emission (dBA) ≦65dBA
Display 7 inches touch screen
UV Timer Yes
Main Power Switch Yes
Alarm Type Audible + Visual
Alarm Function Inflow Velocity, Downflow Velocity, Sash Height Limit, Filter Blocked, Filter Lifetime Remind
Cabinet Full Load Amps (FLA) 2.2A
Power Consumption 1500W
Power Supply AC220±10%,50/60H
Normal Size 5 Fee
Usable Working Area 1.0752m2, 11.57 sq.ft
Stand Height 640mm (Adjustable)
Internal Work Area Dimensions 1600 × 672 × 630mm
External Dimensions with Base Stand 1780 × 790 × 2080mm
Packing Dimension 1910 × 950 × 2220mm
Net Weight 300 Kg
Gross Weight 340 Kg

Features

  • 70% recirculated airflow with 30% top exhaust configuration
  • Microprocessor-regulated airflow stabilization system
  • ULPA filtration ≥99.999% efficiency at 0.1–0.3 µm
  • Average inflow velocity 0.50 m/s with 1277 m³/h downflow volume
  • Type 304 stainless steel work zone with UV interlock safety system

Applications

FunctionImplementation
Aerosol containment549 m³/h exhaust airflow with ULPA filtration barrier
Operator protection validation≤10 CFU impaction sampler performance
Product sterility controlISO 14644.1 Class 3 clean air environment
Cross contamination reduction≤2 CFU cross contamination test result

Standard Accessories

  • Fluorescent lamp – 30W × 2
  • Ultraviolet lamp - 30W × 1
  • Receptacle
  • Detachable Base Stand

Optional Accessories

  • Gas Valve
  • Vacuum Valve
  • Water Valve

Related Models Of Biological Safety Cabinet Class II A2

Lab Expo Biological Safety Cabinet Class II A2 supports controlled containment workflows where personnel, product, and environmental protection are required during biological handling procedures. This biosafety cabinet class 2 operates with HEPA-filtered inflow and downflow airflow to maintain a stable sterile work zone. Digital airflow monitoring manages real-time performance, while integrated alarms and protective air barriers support controlled operation and consistent containment during routine laboratory use.