Biological Safety Cabinet Class II B2 LE-BSC402

Biological Safety Cabinet Class II B2 LE-BSC402

Biological Safety Cabinet Class II B2 LE-BSC402 operates on a 100% exhaust system to eliminate contaminated air completely. It is equipped with a quiet, high-efficiency motor to ensure stable performance under continuous operation. Incorporated airflow sensors detect real-time variations, maintaining secure inflow and downflow conditions. Built-in LCD interface displays critical parameters including pressure, airflow velocity, and system status. Our Biological Safety Cabinet Class II B2 is ideal for environments where recirculation of filtered air is prohibited.

Specifications

Type of BSC Class Ⅱ Type B2
Airflow Mode 100% Exhaust
Exhaust Direction Top Exhaust
Average Inflow Velocity 0.53m/s
Average Downflow Velocity 0.35m/s
Inflow Airflow Volume 1484m3/h
Downflow Airflow Volume 1038m3/h
Exhaust Airflow Volume 1484m3/h
Velocity Meter Accuracy 0.001m/s
Cleanliness ULPA: Class 3(ISO 14644.1); Class 100 (Fed 209E)
Filter Typical Efficiency Greater Than 99.999% at 0.1 to 0.3 Micron, ULPA as per IEST-RP-CC001.3 USA
Total Colony in Impaction Sampler (Operator Protection) ≤10CFU/Time
Total Colony in Slit Type Sampler (Operator Protection) ≤5CFU/Time
Total Colony in Culture Dish (Product Protection) ≤5CFU/Time
Total Colony in Culture Dish (Cross Contamination) ≤2CFU/Time
Personnel Protection Test KI-Discus Containment and Microbiological Testing are Performed
Blower AC
Downflow Filter ULPA
Exhaust Filter ULPA
Main Body 1.2mm (xxx'') 18 Gauge Electro-Galvanized Steel with White Oven-Baked Powder-Coated Finish
Work Zone 1.5mm (xxx'') 16 Gauge Stainless Steel, Type304, Single-Piece or 3-Piece
Side Walls 1.5mm (Xxx'') 16 Gauge Stainless Steel, Type304
Sash Glass Thickness 5mm
Sash Glass Type Tempered Glass, UV-Proof
Sash Window Working Opening 180mm
Sash Maximum Opening 530mm
Fluorescent Light Intensity ≥700lux
RMS ≤3um
Sound Emission (dBA) ≦65dBA
Display 7 Inches Touch Screen
UV Timer Yes
Main Power Switch Yes
Alarm Type Audible + Visual
Alarm Function Inflow Velocity, Downflow Velocity, Sash Height Limit, Filter Blocked, Filter Lifetime Remind
Cabinet Full Load Amps (FLA) 2.2A
Power Consumption 1500W (Excluding External Exhaust Fan)
Power Supply AC220±10%,50/60H
Normal Size 4 Feet
Usable Working Area 0.8736m2, 9.40 sq.ft
Stand Height 640mm (Adjustable)
Internal Work Area Dimensions 1300 × 672 × 630mm
External Dimensions with Base Stand 1480 × 790 × 2080mm
Packing Dimension 1610 × 950 × 2220mm
Net Weight 300 Kg
Gross Weight 330 Kg

Features

  • Class II Type B2 cabinet with 100% external exhaust airflow
  • Average inflow velocity 0.53 m/s with 1484 m³/h exhaust volume
  • Downflow airflow volume 1038 m³/h with ULPA filtration ≥99.999% at 0.1–0.3 µm
  • Air velocity monitoring accuracy 1×10⁻³ m/s with real-time LCD parameter display

Applications

FunctionImplementation
Complete contaminated air removal100% top exhaust ducted configuration
Personnel protection validation≤10 CFU impaction sampler containment test
Product sterility controlISO 14644.1 Class 3 vertical downflow environment
Cross contamination restriction≤2 CFU cross contamination performance
Non-recirculating hazardous proceduresFully ducted cabinet for prohibited recirculation environments

Standard Accessories

  • Fluorescent lamp – 30W × 2
  • Ultraviolet lamp - 30W × 1
  • Receptacle
  • Detachable Base Stand
  • Wheels

Optional Accessories

  • Gas Valve
  • Vacuum Valve
  • Water Valve

Biological Safety Cabinet Class II B2 LE-BSC402

Biological Safety Cabinet Class II B2 LE-BSC402


Related Models Of Biological Safety Cabinet Class II B2

Lab Expo Biological Safety Cabinet Class II B2 supports high-containment laboratory operations where total exhaust and zero air recirculation are required for hazardous biological and chemical handling. This bsc class ii type b2 operates with full exhaust airflow and HEPA filtration to remove contaminated air from the work zone. Digital controls monitor airflow and pressure, while integrated alarms and safety interlocks ensure controlled, compliant containment during critical procedures.