Biological Safety Cabinet Class II A2 LE-BSC302 Catalog

Download

Overview :

Biological Safety Cabinet Class II A2 LE-BSC302 delivers 70% internal air recirculation and 30% exhaust, ensuring optimal containment and sterile working conditions. It integrates a microprocessor-based airflow control system that auto-adjusts volume without manual intervention. Equipped with overload protection, it supports reliable performance. Integrated large LCD screen and intuitive interface provide real-time monitoring, enhancing safety and operational ease. Our Biological Safety Cabinet Class II A2 is used for sterile procedures involving pathogenic microorganisms or sensitive biological materials.

Features:

  • 70% recirculated airflow with 30% top exhaust configuration
  • ULPA filtration efficiency ≥99.999% at 0.1–0.3 µm particle size
  • Average inflow velocity 0.50 m/s with downflow velocity 0.33 m/s
  • Airflow volumes: 446 m³/h inflow, 1038 m³/h downflow, 446 m³/h exhaust
  • Velocity meter resolution 1×10⁻³ m/s with 7-inch touch screen monitoring

Applications:

FunctionImplementationAerosol containmentKI-Discus tested airflow barrier systemSterile processingISO Class 3 (ULPA) vertical laminar downflow environmentMicrobial load control≤5 CFU product protection performanceCross contamination control≤2 CFU validated separation performancePathogen handlingEnclosed stainless steel work zone for biological materials

Specifications:
Type of BSC Class Ⅱ Type A2
Airflow Mode 70% Cycle, 30% Exhaust
Exhaust Direction Top Exhaust
Average Inflow Velocity 0.50m/s
Average Downflow Velocity 0.33m/s
Inflow Airflow Volume 446m3/h
Downflow Airflow Volume 1038m3/h
Exhaust Airflow Volume 446m3/h
Velocity Meter Accuracy 0.001m/s
Cleanliness ULPA: Class 3(ISO 14644.1); Class 100 (Fed 209E)
Filter Typical Efficiency Greater Than 99.999% at 0.1 to 0.3 Micron, ULPA as per IEST-RP-CC001.3 USA
Total Colony in Impaction Sampler (Operator Protection) ≤10CFU/Time
Total Colony in Slit Type Sampler (Operator Protection) ≤5CFU/Time
Total Colony in Culture Dish (Product Protection) ≤5CFU/Time
Total Colony in Culture Dish (Cross Contamination) ≤2CFU/Time
Personnel Protection Test KI-Discus Containment and Microbiological Testing are Performed
Blower AC
Downflow Filter ULPA
Exhaust Filter ULPA
Main Body 1.2mm (xxx'') 18 Gauge Electro-Galvanized Steel with White Oven-Baked Powder-Coated Finish
Work Zone 1.5mm (xxx'') 16 Gauge Stainless Steel, Type304, Single-Piece or 3-Piece
Side Walls 1.5mm (Xxx'') 16 Gauge Stainless Steel, Type304
Sash Glass Thickness 5mm
Sash Glass Type Tempered Glass, UV-Proof
Sash Window Working Opening 180mm
Sash Maximum Opening 530mm
Fluorescent Light Intensity ≥700lux
RMS ≤3um
Sound Emission (dBA) ≦65dBA
Dis, play 7 inches touch screen
UV Timer Yes
Main Power Switch Yes
Alarm Type Audible + Visual
Alarm Function Inflow Velocity, Downflow Velocity, Sash Height Limit, Filter Blocked, Filter Lifetime Remind
Cabinet Full Load Amps (FLA) 2.2A
Power Consumption 1500W
Power Supply AC220±10%,50/60H
Normal Size 4 Feet
Usable Working Area 0.8736m2, 9.40 sq.ft.
Stand Height 640mm (Adjustable)
Internal Work Area Dimensions 1300 × 672 × 630mm
External Dimensions with Base Stand 1480 × 790 × 2080mm
Packing Dimension 1610 × 950 × 2220mm
Net Weight 270 Kg
Gross Weight 300 Kg
Standard Accessories:
  • Fluorescent lamp – 30W × 2
  • Ultraviolet lamp - 30W × 1
  • Receptacle
  • Detachable Base Stand
Optional Accessories:
  • Gas Valve
  • Vacuum Valve
  • Water Valve
www.labexpo.com