Biological Safety Cabinet Class II A2 LE-BSC302

Biological Safety Cabinet Class II A2 LE-BSC302

Biological Safety Cabinet Class II A2 LE-BSC302 delivers 70% internal air recirculation and 30% exhaust, ensuring optimal containment and sterile working conditions. It integrates a microprocessor-based airflow control system that auto-adjusts volume without manual intervention. Equipped with overload protection, it supports reliable performance. Integrated large LCD screen and intuitive interface provide real-time monitoring, enhancing safety and operational ease. Our Biological Safety Cabinet Class II A2 is used for sterile procedures involving pathogenic microorganisms or sensitive biological materials.

Specifications

Type of BSC Class Ⅱ Type A2
Airflow Mode 70% Cycle, 30% Exhaust
Exhaust Direction Top Exhaust
Average Inflow Velocity 0.50m/s
Average Downflow Velocity 0.33m/s
Inflow Airflow Volume 446m3/h
Downflow Airflow Volume 1038m3/h
Exhaust Airflow Volume 446m3/h
Velocity Meter Accuracy 0.001m/s
Cleanliness ULPA: Class 3(ISO 14644.1); Class 100 (Fed 209E)
Filter Typical Efficiency Greater Than 99.999% at 0.1 to 0.3 Micron, ULPA as per IEST-RP-CC001.3 USA
Total Colony in Impaction Sampler (Operator Protection) ≤10CFU/Time
Total Colony in Slit Type Sampler (Operator Protection) ≤5CFU/Time
Total Colony in Culture Dish (Product Protection) ≤5CFU/Time
Total Colony in Culture Dish (Cross Contamination) ≤2CFU/Time
Personnel Protection Test KI-Discus Containment and Microbiological Testing are Performed
Blower AC
Downflow Filter ULPA
Exhaust Filter ULPA
Main Body 1.2mm (xxx'') 18 Gauge Electro-Galvanized Steel with White Oven-Baked Powder-Coated Finish
Work Zone 1.5mm (xxx'') 16 Gauge Stainless Steel, Type304, Single-Piece or 3-Piece
Side Walls 1.5mm (Xxx'') 16 Gauge Stainless Steel, Type304
Sash Glass Thickness 5mm
Sash Glass Type Tempered Glass, UV-Proof
Sash Window Working Opening 180mm
Sash Maximum Opening 530mm
Fluorescent Light Intensity ≥700lux
RMS ≤3um
Sound Emission (dBA) ≦65dBA
Dis, play 7 inches touch screen
UV Timer Yes
Main Power Switch Yes
Alarm Type Audible + Visual
Alarm Function Inflow Velocity, Downflow Velocity, Sash Height Limit, Filter Blocked, Filter Lifetime Remind
Cabinet Full Load Amps (FLA) 2.2A
Power Consumption 1500W
Power Supply AC220±10%,50/60H
Normal Size 4 Feet
Usable Working Area 0.8736m2, 9.40 sq.ft.
Stand Height 640mm (Adjustable)
Internal Work Area Dimensions 1300 × 672 × 630mm
External Dimensions with Base Stand 1480 × 790 × 2080mm
Packing Dimension 1610 × 950 × 2220mm
Net Weight 270 Kg
Gross Weight 300 Kg

Features

  • 70% recirculated airflow with 30% top exhaust configuration
  • ULPA filtration efficiency ≥99.999% at 0.1–0.3 µm particle size
  • Average inflow velocity 0.50 m/s with downflow velocity 0.33 m/s
  • Airflow volumes: 446 m³/h inflow, 1038 m³/h downflow, 446 m³/h exhaust
  • Velocity meter resolution 1×10⁻³ m/s with 7-inch touch screen monitoring

Applications

FunctionImplementation
Aerosol containmentKI-Discus tested airflow barrier system
Sterile processingISO Class 3 (ULPA) vertical laminar downflow environment
Microbial load control≤5 CFU product protection performance
Cross contamination control≤2 CFU validated separation performance
Pathogen handlingEnclosed stainless steel work zone for biological materials

Standard Accessories

  • Fluorescent lamp – 30W × 2
  • Ultraviolet lamp - 30W × 1
  • Receptacle
  • Detachable Base Stand

Optional Accessories

  • Gas Valve
  • Vacuum Valve
  • Water Valve

Related Models Of Biological Safety Cabinet Class II A2

Lab Expo Biological Safety Cabinet Class II A2 supports controlled containment workflows where personnel, product, and environmental protection are required during biological handling procedures. This biosafety cabinet class 2 operates with HEPA-filtered inflow and downflow airflow to maintain a stable sterile work zone. Digital airflow monitoring manages real-time performance, while integrated alarms and protective air barriers support controlled operation and consistent containment during routine laboratory use.