Preparative HPLC 37-PHP100 Catalog

Download

Overview :

Preparative HPLC is equipped with 4 pump head variation and a UV or RI detector with a wavelength range of 190 to 400 nm and 400 to 800 nm. Built in multi positioned fraction valve and controlled fractionation optimizes controlled fraction collection and solvent recycling. With a column size of 10 x 1000 mm the retention, resolution and selectivity is optimized.

Features:

  • Solvent flow rate is 500 ml / min
  • Multiple chromatogram view
  • 27 integrated programmable parameters like peak width, tangent slope etc
  • Compatibility for medium and low pressure application
  • 10 points pressure calibration for pressure linearity
  • Auto apply of mathematical operations to chromatograms
  • Automated sequential commands post batch run
  • External pump control by panel or computer software
  • Addition of tools for detector controlled fraction collection and solvent recycling
  • Controlled fractionation with option of manual control during the run
  • Error alarm function for over pressure ,under pressure and leakage

Applications:

Used for purification of peptides and proteins, separation of high molecular mass compounds, separation of protein mixtures, qualitative and quantitative determination of a compound, automated fraction re-analysis

Specifications:
Dimension 750 x 860 x 1280 mm ( 100 ml , 250 ml , 500 ml , 1000 ml ) , 810 x 950 x 1310 mm ( 3 L )
Wetted material 316 L and PTFE
Power < 1500 W
Mobility With castor
Pump Specifications:
Model No SHP-A1
Pump head capacity with max pressure 100 ml ; 30 MPa
250 ml ; 50 MPa
500 ml ; 10 MPa
1000 ml ; 10 MPa
3000 ml ; 10 MPa
Pressure pulsation 1 % at 10 MPa
Flow rate accuracy ± 1 %
Flow rate reproducibility ≤ 0.5
Gradient accuracy ± 2 %
Pump inlet tube ᴓ 4 ( 100 ml , 250 ml , 500 ml )
ᴓ 6 ( 1000 ml )
ᴓ 10 ( 3000 ml )
Pump outlet tube 1/8 th ( 100 ml , 250 ml , 500 ml )
DN4 ( 1000 ml )
DN6 ( 3000 ml )
Detector Specifications:
Model No SHP-B1
Wavelength range 190 ~ 400 nm / 400 ~ 800 nm
Wavelength accuracy ± 2 nm
Wavelength reproducibility 0.4 nm
Bandwidth 8 nm
Baseline noise ( Au ) 3 x 10 -5
Baseline draft ( Au / h ) 1 x 10 -4
Optical path 10 mm
Flow cell PEEK
Fraction Valve Specifications:
Model No SHP-C1
Valve type Ball-valve ( 8 to 16 port )
Valve head material Stainless steel / PEEK
Capillary connection 1” / 4”
Maximum pressure stability 10 bar
Driver mode Air or electric
www.labexpo.com